ArtikelThe effect of calibration and detector temperature on the reconstructed cone beam computed tomography image quality: a study for the workflow of the iCAT Classic equipment  
Oleh: Plachtovics, Mark ; Goczan, Janos ; Nagy, Katalin
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology (keterangan: ada di ClinicalKey) vol. 119 no. 04 (Apr. 2015), page 473–480 .
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: O04.K
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelObjective The image quality of the reconstructed dental cone beam computed tomography (CBCT) acquisition is strongly dependent on the characteristics of the applied detector. Some flat panel detector (FPD) types require a warming-up period (WUP) to achieve a steady-state temperature; and some of them, those that have a cesium iodide detector, may require a calibration process (CP) after each WUP. Study Design Quality-assurance phantom was used for the evaluation of image quality, including spatial accuracy and density response with and without WUP and CP using iCAT Classic equipment. Results The correlation between the measured gray values and the multidetector computed tomography (MDCT) values deviated slightly from linearity (defined by the 45-degree line). There was no detectable difference in the spatial accuracy of the four different scanning modes. Conclusions Although the WUP is important to reach the required steady-state temperature, the CP has a greater effect on the image quality.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0.03125 second(s)