ArtikelImproved Optical Probe for Measuring Phytoplankton Suspension Concentrations Based on Optical Fluoresensce and Absorption  
Oleh: Purnamaningsih, Retno Wigajatri ; Prihantini, Nining Betawati
Jenis: Article from Journal - ilmiah nasional - terakreditasi DIKTI
Dalam koleksi: Makara Journal of Technology vol. 16 no. 2 (Nov. 2012), page 116-120 .
Topik: Optical Probe; Optical Fluorescence; Optical Absorption; Phytoplankton Concentration
Fulltext: 241-271-1-PB.pdf (290.54KB)
Ketersediaan
  • Perpustakaan Pusat (Semanggi)
    • Nomor Panggil: MM64
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelThis paper discusses the results of improved optical probe that works according to optical fluorescence and absorption phenomena for measuring the phytoplankton suspension concentrations. Measurements are made on the Scenedesmus sp. culture. The laboratory test has shown that range of concentration from 104 up to around 1 x 106 cells/mL; fluorescence intensity at ?= 685 nm; and logarithmic of transmission intensity at ?= 405 nm are proportional to the cell concentration linear with proportional constant ? = 4 × 10-5 and ß = –2 × 10-7 mL/cell respectively.