ArtikelComparability of antimullerian hormone levels among commercially available immunoassays  
Oleh: Su, H. Irene ; Sammel, Mary D. ; Homer, Michael V. ; Kim, Bui ; Haunschild, Carolyn ; Stanczyk, Frank Z.
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: Fertility and Sterility (keterangan: ada di ClinicalKey) vol. 101 no. 06 (Jun. 2014), page 1766–1772.
Topik: AMH; immunoassay; breast cancer
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: F02.K.2014.02
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelObjective To compare antimüllerian hormone (AMH) levels among three commercially available AMH immunoassays: AMH Gen II (Beckman Coulter), Ultrasensitive AMH (Ansh Labs), and picoAMH (Ansh Labs). Design Cross-sectional. Setting Academic reproductive endocrinology program. Patient(s) 90 newly diagnosed breast cancer patients before cancer treatment. Intervention(s) None. Main Outcome Measure(s) Proportion of detectable AMH levels by immunoassay, and comparability among assays. Result(s) At a mean age of 38.1 years, the median (interquartile range) AMH level for the cohort was 0.92 [1.35] ng/mL for the Gen II assay, 1.68 [2.30] ng/mL for the Ultrasensitive assay, and 1.52 [2.41] ng/mL for the picoAMH assay. Significantly higher proportions of detectable AMH levels were observed with the picoAMH kit (97%) compared with both the Gen II (84%) and Ultrasensitive (92%) assays. Although the AMH results were highly correlated among the assays (r = 0.92–0.99), the Gen II AMH levels were consistently lower than both Ultrasensitive and picoAMH levels. Moreover, as AMH levels increased, the magnitude of difference grew larger between Gen II and each of the other two assays. Conclusion(s) Measurement of AMH levels with the picoAMH kit maximized detection at very low levels, particularly in contrast with the Gen II kit. Conversion of AMH levels from different immunoassays using regression equations is potentially highly inaccurate.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0.015625 second(s)