ArtikelKalibrasi Sensor Kawat Panas Anemoneter Pada Temperatur Konstan  
Oleh: Ruslan, Wegie
Jenis: Article from Journal - ilmiah nasional - tidak terakreditasi DIKTI - atma jaya
Dalam koleksi: Atma nan Jaya vol. 5 no. 3 (Dec. 1992), page 1-14.
Topik: ANEMOMETER; kalibrasi sensor kawat panas; anemoneter; temperatur konstan
Fulltext: Wegie Ruslan.pdf (646.37KB)
Ketersediaan
  • Perpustakaan Pusat (Semanggi)
    • Nomor Panggil: AA48.3
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
  • Perpustakaan PKPM
    • Nomor Panggil: A66
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelCalibration of a hot wire pole anemometer is to be completed prior to the measurement of velocity of turbulence. This calibration is carried outwith the assumption that the ambient temperature is constant. using King's and Bernoulli's Laws we obtain the formulas of velocity and turbulence which contain the constants n and B (with the potential difference) as the variable that can be measured by a voltmeter. The relation between velocity and potential difference is illustrated in the form of the linear relation between log U and log (E2 - E2 O). This relation yield the constants n and B which can be used in calculating either velocity or turbulence.