Anda belum login :: 16 Apr 2025 10:06 WIB
Home
|
Logon
Hidden
»
Administration
»
Collection Detail
Detail
Pressure changes during embryo transfer
Oleh:
Grygoruk, Cezary
;
Sieczynski, Piotr
;
Pietrewicz, Piotr
;
Mrugacz, Malgorzata
;
Gagan, Jerzy
;
Mrugacz, Grzegorz
Jenis:
Article from Journal - ilmiah internasional
Dalam koleksi:
Fertility and Sterility (keterangan: ada di ClinicalKey) vol. 95 no. 02 (Feb. 2011)
,
page 538-541.
Topik:
FERTILITY
;
Catheter loading
;
ejection speed
;
embryo transfer
;
pressure injury
Ketersediaan
Perpustakaan FK
Nomor Panggil:
F02.K.2011.02
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
Objective To investigate the pressure changes in the transferred load during mock ET. Design Experimental setup. Setting Academic Research Institute of Mechanical Engineering and private centers of reproductive medicine. Patients(s) None. Intervention(s) None. Main Outcome Measure(s) Laboratory simulations of ET into a rigid transparent uterine model equipped with a pressure sensor. Result(s) Injection of a transferring load during mock ET could increase pressure locally up to 155 mm Hg in <0.1 seconds. The recorded pressure increase slope reached values as high as 72,000 mmHg/s, and the pressure decrease slope reached 144,000 mmHg/s. The pressure buildup in the transferred liquid was proportional to the ejection speed of the transferred load. Conclusion(s) ET can cause rapid pressure fluctuations in the transferred liquid. Therefore, it is advisable to transfer the embryo gently with minimum ejection speed, to avoid exposing the embryo to the steep pressure gradient.
Opini Anda
Klik untuk menuliskan opini Anda tentang koleksi ini!
Kembali
Process time: 0 second(s)