Anda belum login :: 16 Apr 2025 19:54 WIB
Detail
ArtikelReview of Adverse Events Associated With False Glucose Readings Measured by GDH-PQQ–Based Glucose Test Strips in the Presence of Interfering Sugars  
Oleh: Frias, Juan P. ; Lim, Christine G. ; Ellison, John M.
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: Diabetes Care vol. 33 no. 04 (Apr. 2010), page 728-729 .
Topik: glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ)
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: D05.K.2010.02
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelOBJECTIVE To assess the implications of falsely elevated glucose readings measured with glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ) test strips. RESEARCH DESIGN AND METHODS We conducted a review of the Food and Drug Administration's Manufacturer and User Facility Device Experience database and medical literature for adverse events (AEs) associated with falsely elevated glucose readings with GDH-PQQ test strips in the presence of interfering sugars. RESULTS Eighty-two reports were identified: 16 (20%) were associated with death, 46 (56%) with severe hypoglycemia, and 12 (15%) with nonsevere hypoglycemia. In eight reports (10%), the AE was not described. Forty-two events (51%) occurred in the U.S. Although most events occurred in hospitalized patients, at least 14 (17%) occurred in outpatients. Agents most commonly associated with AEs were icodextrin-containing peritoneal dialysate and maltose-containing intravenous immune globulin. CONCLUSIONS GDH-PQQ test strips pose a safety risk to insulin-using patients treated with agents containing or metabolized to interfering sugars.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0.015625 second(s)