Isi artikelOBJECTIVE The potential benefits of continuous glucose monitoring (CGM) in the management of adults and children with well-controlled type 1 diabetes have not been examined. RESEARCH DESIGN AND METHODS A total of 129 adults and children with intensively treated type 1 diabetes (age range 8–69 years) and A1C <7.0% were randomly assigned to either continuous or standard glucose monitoring for 26 weeks. The main study outcomes were time with glucose level =70 mg/dl, A1C level, and severe hypoglycemic events. RESULTS At 26 weeks, biochemical hypoglycemia (=70 mg/dl) was less frequent in the CGM group than in the control group (median 54 vs. 91 min/day), but the difference was not statistically significant (P = 0.16). Median time with a glucose level =60 mg/dl was 18 versus 35 min/day, respectively (P = 0.05). Time out of range (=70 or >180 mg/dl) was significantly lower in the CGM group than in the control group (377 vs. 491 min/day, P = 0.003). There was a significant treatment group difference favoring the CGM group in mean A1C at 26 weeks adjusted for baseline (P < 0.001). One or more severe hypoglycemic events occurred in 10 and 11% of the two groups, respectively (P = 1.0). Four outcome measures combining A1C and hypoglycemia data favored the CGM group in comparison with the control group (P < 0.001, 0.007, 0.005, and 0.003). CONCLUSIONS Most outcomes, including those combining A1C and hypoglycemia, favored the CGM group. The weight of evidence suggests that CGM is beneficial for individuals with type 1 diabetes who have already achieved excellent control with A1C <7.0%.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0 second(s)