A Novel Technique to Measure The Magnitude and Direction of Flow in A Tube
Oleh:
Robinson, Allen L.
;
Sextro, Richard G.
Jenis:
Article from Bulletin/Magazine
Dalam koleksi:
Journal of Fluids Engineering vol. 122 no. 1 (2000)
,
page 186-188.
Topik:
TUBES
;
novel technique
;
magnitude
;
direction
;
flow
;
tube
Ketersediaan
Perpustakaan Pusat (Semanggi)
Nomor Panggil:
JJ89.1
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
This paper describes a novel in - line sensor that measures the magnitude and direction of gas flow in a tube. The sensor possesses a unique set of performance characteristics : low detection limit, little resistance to flow, and directional sensitivity. The sensor consists of two hot wire anemometers mounted in a U - shaped tube. Differences in the signals between the two hot wires under low velocity conditions are used to determine the direction of the flow. Calibration curves of flow rate versus measured velocity are used to determine the magnitude of the flow. The sensor has applications in systems that are characterized by naturally driven oscillating flows.