ArtikelCentral Corneal Thickness and Corneal Hysteresis During Corneal Swelling Induced by Contact Lens Wear With Eye Closure  
Oleh: Lu, Fan ; Xu, Suzhong ; Qu, Jia ; Shen, Meixiao ; Wang, Xiaoxing ; Fang, Haizhen ; Wang, Jianhua
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: American Journal of Ophthalmology (keterangan: ada di ClinicalKey) vol. 143 no. 04 (Apr. 2007), page 616.
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: A12.K.2007.01
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelPurpose To determine if corneal hysteresis (CH) was associated with increased central corneal thickness (CCT) induced by wearing soft contact lenses during eye closure. Design A prospective laboratory investigation. Methods CCT was measured with a modified optical coherence tomography (OCT), and CH was measured with a Reichert Ocular Response Analyzer (ORA) [Reichert Ophthalmic Instruments, Depew, New York, USA]. The ORA also determined values for intraocular pressure (IOP), corneal compensated IOP (IOPcc), and corneal resistance factor (CRF). One randomly selected eye of 20 non–contact lens wearers (four males and 16 females, age 19.7 ± 1.1 years) was patched during three hours of soft contact lens wear. Measurements were made before lens insertion, immediately upon removal, and every 20 minutes thereafter for 100 minutes. Results Immediately after contact lens removal, CCT was increased by 13.1 ± 2.2% (mean ± SD) compared with baseline (post hoc, P = .001). After 100 minutes, it remained elevated by 2.4 ± 1.6% (post hoc, P = .001). However, there were no significant differences of CH at any time after lens wear (analysis of variance [ANOVA], P = .9). Immediately after lens removal, there were significant increases in IOP (post hoc, P = .003) and corneal resistance factor (CRF) (post-hoc, P = .015), but not in IOPcc (post hoc, P = .07). After lens wear, there were significant but weak correlations between the percentage change of CCT (CCT%) and IOP (r = 0.32, P = .001) and IOPcc (r = 0.29, P = .001). However, there was no significant correlation between CCT% and CH (r = 0.07, P = .458). Conclusion CH as measured by ORA was not associated with corneal swelling induced by soft contact lens wear in this study group.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0 second(s)