Design Considerations for Artificial Pancreas Pivotal Studies
Oleh:
Russell, Steven J.
;
Beck, Roy W.
Jenis:
Article from Journal - ilmiah internasional
Dalam koleksi:
Diabetes Care vol. 39 no. 07 (Jul. 2016)
,
page 1161-1167.
Topik:
Artificial Pancreas
;
AP
Ketersediaan
Perpustakaan FK
Nomor Panggil:
D05.K
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
The development of artificial pancreas systems has evolved to the point that pivotal studies designed to assess efficacy and safety are in progress or soon to be initiated. These pivotal studies are intended to provide the necessary data to gain clearance from the U.S. Food and Drug Administration, coverage by payers, and adoption by patients and clinicians. Although there will not be one design that is appropriate for every system, there are certain aspects of protocol design that will be considerations in all pivotal studies designed to assess efficacy and safety. One key aspect of study design is the intervention to be used by the control group. A case can be made that the control group should use the currently available best technology, which is sensor-augmented pump therapy. However, an equally, if not more, compelling case can be made that the control intervention should be usual care. In this Perspective, we elaborate on this issue and provide a pragmatic approach to the design of clinical trials of artificial pancreas systems.