Anda belum login :: 27 Nov 2024 03:07 WIB
Home
|
Logon
Hidden
»
Administration
»
Collection Detail
Detail
Effects of Stem Design and Pre-cooling Prostheses on the Heat Generated by Bone Cement in an In vitro Model
Oleh:
Rodop, O
;
Kiral, A
;
Arpacioglu, O
Jenis:
Article from Journal - ilmiah internasional
Dalam koleksi:
The Journal of International Medical Research vol. 30 no. 03 (May 2002)
,
page 265-270.
Topik:
polymethylmethacrylate
;
bone cement
;
thermal effect
;
in vitro model
;
prostheses
Ketersediaan
Perpustakaan FK
Nomor Panggil:
J11.K
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
The necrotizing effects of the heat, particularly at more than 50°C, produced by the exothermic polymerization process associated with the acrylic implant cement polymethylmethacrylate (PMMA) are well documented. The temperature changes that occur are dependent on the thickness of the PMMA. The current study investigates the hypothesis that the heat produced by the bone cement may be reduced by the choice of stem design and by pre-cooling the hip prosthesis. The thermal alterations at the bone–cement interface were measured in an in vitro model. The results indicated that a temperature decrease of approximately 7°C could be achieved by pre-cooling the prosthesis, and by changing the shape of the prosthesis stem from flat and wide to round.