Anda belum login :: 23 Nov 2024 09:48 WIB
Home
|
Logon
Hidden
»
Administration
»
Collection Detail
Detail
PENENTUAN KADAR LIPOPROTEIN RAPATAN TINGGI (HIGH DENSITY) DENGAN DUA PEREAKSI (REAGEN) BERBEDA MENGGUNAKAN HITACHI 902
Oleh:
Hutagalung, J.
;
Arif, Mansyur
Jenis:
Article from Journal - ilmiah nasional - terakreditasi DIKTI
Dalam koleksi:
Indonesian Journal of Clinical Pathology and Medical Laboratory vol. 15 no. 02 (Mar. 2009)
,
page 46-48.
Topik:
HDL
;
Hitachi 902
;
Daichi reagent
;
Roche reagent
Ketersediaan
Perpustakaan FK
Nomor Panggil:
I01.K.02
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
Reagent selection is one of the factors that could influence the quality of laboratory results. The use of open system tools gives the possibility to choose the best reagents, including the reagent for high density lipoprotein (HDL) determination. The aim of this study was to compare HDL level determination using two different reagents measured by Hitachi 902. A cross sectional study was done from January to February 2007 in Ratulangi Medical Centre Laboratory, Makassar. From 47 samples we found that the mean HDL level using Daichi reagent was 50.47 mg/dl ranging from 45.99 mg/dl to 54.94 mg/dl and the mean using Roche reagent was 56.23 mg/dl ranging from 50.93 mg/dl to 61.53 mg/dl with p = 0.098, and Pearson Correlation was 0.900 with p = 0.000. There was no significant difference between HDL level measured by Hitachi 902 using Daichi and Roche reagents.
Opini Anda
Klik untuk menuliskan opini Anda tentang koleksi ini!
Kembali
Process time: 0.015625 second(s)