Anda belum login :: 27 Nov 2024 09:17 WIB
Home
|
Logon
Hidden
»
Administration
»
Collection Detail
Detail
Lesion Detectability and Clinical Effectiveness of Dual-head Coincidence Gamma Camera Imaging in Comparison with Dedicated PET Systems in Tumour Patients
Oleh:
Narin, Y.
;
Urhan, M.
;
Canpolat, N.
;
Vardereli, E.
;
Bayhan, H.
Jenis:
Article from Journal - ilmiah internasional
Dalam koleksi:
The Journal of International Medical Research vol. 35 no. 04 (Jul. 2007)
,
page 467-473.
Topik:
Gamma Camera Imaging
;
PET Systems
;
tumour
Ketersediaan
Perpustakaan FK
Nomor Panggil:
J11.K.2007.01
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
The lesion detection capability and clinical effectiveness of dual-head coincidence gamma camera imaging (c-PET) were compared with those of dedicated positron emission tomography (d-PET) in 37 cancer patients who underwent whole-body c-PET and d-PET imaging after administration of 370 - 540 MBq 18F-fluorodeoxyglucose. Eighty-nine lesions were detected on c-PET whereas 133 lesions were seen with d-PET imaging. The relative sensitivity of c-PET compared with d-PET was 62% and 73% for lesions < 15 and >/= 15 mm, respectively, and the relative concordance rate was 84% when the patients were restaged. Since the lesion detection rate of c-PET imaging was lower than that of d-PET, the detection of small lesions, therefore, requires care. The clinical effectiveness of c-PET, however, was similar to that of d-PET and, therefore, it is concluded that c-PET can be used as an alternative to d-PET, particularly considering the high cost and limited availability of d-PET cameras.
Opini Anda
Klik untuk menuliskan opini Anda tentang koleksi ini!
Kembali
Process time: 0.015625 second(s)