Anda belum login :: 27 Nov 2024 09:17 WIB
Detail
ArtikelLesion Detectability and Clinical Effectiveness of Dual-head Coincidence Gamma Camera Imaging in Comparison with Dedicated PET Systems in Tumour Patients  
Oleh: Narin, Y. ; Urhan, M. ; Canpolat, N. ; Vardereli, E. ; Bayhan, H.
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: The Journal of International Medical Research vol. 35 no. 04 (Jul. 2007), page 467-473.
Topik: Gamma Camera Imaging; PET Systems; tumour
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: J11.K.2007.01
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelThe lesion detection capability and clinical effectiveness of dual-head coincidence gamma camera imaging (c-PET) were compared with those of dedicated positron emission tomography (d-PET) in 37 cancer patients who underwent whole-body c-PET and d-PET imaging after administration of 370 - 540 MBq 18F-fluorodeoxyglucose. Eighty-nine lesions were detected on c-PET whereas 133 lesions were seen with d-PET imaging. The relative sensitivity of c-PET compared with d-PET was 62% and 73% for lesions < 15 and >/= 15 mm, respectively, and the relative concordance rate was 84% when the patients were restaged. Since the lesion detection rate of c-PET imaging was lower than that of d-PET, the detection of small lesions, therefore, requires care. The clinical effectiveness of c-PET, however, was similar to that of d-PET and, therefore, it is concluded that c-PET can be used as an alternative to d-PET, particularly considering the high cost and limited availability of d-PET cameras.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 