Anda belum login :: 27 Nov 2024 08:10 WIB
Detail
ArtikelCardiac Output Measurement by Arterial Waveform Analysis in Cardiac Surgery – a Comparison of Measurements Derived from Waveforms of the Radial Artery versus the Ascending Aorta  
Oleh: Sander, M. ; Spies, C.D. ; Foer, A. ; Heymann, C. von
Jenis: Article from Journal - ilmiah internasional
Dalam koleksi: The Journal of International Medical Research vol. 36 no. 03 (May 2008), page 414-419.
Ketersediaan
  • Perpustakaan FK
    • Nomor Panggil: J11.K.2008.01
    • Non-tandon: 1 (dapat dipinjam: 0)
    • Tandon: tidak ada
    Lihat Detail Induk
Isi artikelIn cardiac surgery, perioperative haemodynamic management is often guided by cardiac output (CO) measurements. The VigileoTM monitor offers uncalibrated CO measurement by arterial waveform analysis (COwave). This validation study compared CO measurements derived from radial artery waveform analysis with those derived from the ascending aorta. CO measurements from the radial artery versus the ascending aorta showed a significant correlation before and after cardiopulmonary bypass (CPB). However, Bland–Altman analysis showed a mean bias of 0.1 l/min and 0.1 l/min, and limits of agreement (LOA) of +2.2 l/min and –1.9 l/min prior to CPB, and +2.5 l/min and –2.7 l/min after weaning from CPB. A comparison of these CO measurements showed a low mean bias, but wide LOA before and after CPB. Therefore measurements using uncalibrated COwave have to be interpreted with caution in a clinical situation.
Opini AndaKlik untuk menuliskan opini Anda tentang koleksi ini!

Kembali
design
 
Process time: 0.03125 second(s)