Desain dan Konstruksi Amf;ifier Spektroskopi Dengan Detektor CsI (TI) Untuk Spektroskopi Nuklir (Gamma)
Oleh:
Setyadi W.S.
;
Isaris, Rill
Jenis:
Article from Journal - ilmiah nasional - tidak terakreditasi DIKTI
Dalam koleksi:
SIGMA: Jurnal Sains dan teknologi vol. 11 no. 2 (Jul. 2008)
,
page 115-126.
Topik:
Design of Amplifier
;
Nuclear (Gamma)
;
Spectroscopy
Ketersediaan
Perpustakaan Pusat (Semanggi)
Nomor Panggil:
SS25.6
Non-tandon:
1 (dapat dipinjam: 0)
Tandon:
tidak ada
Lihat Detail Induk
Isi artikel
New techology on Nuclear (Gamma) Spectroscopy System for element analyis and nuclear experiment has been developed using new Cs(TI) detector, as one of the important component on spectroscopy using CsI(TI) detector, has been designed and constructed. The detector output is 10mV at gamma energy of 662 keV. Design of to nuclear spectroscopy rules. The performance test show that the system linearity is 0.001%, frequency respond up to to 65 kHz, the rise time of 3uS, fall time of 5 uS, shaping time (t) is 9 uS and Full Width Half Maximum (FWHM) is 7 uS. Dyanamic test result using Cs(TI) detector shows 10.2% resolution. These results are satisfactory for nuclear spectroscopy amplifier. The application of Work BEnch Program takes important part in simulation and optimizes the lectronic circuit.